IM 12J05D02-01E
Parameter setting 5-21
Code Display Function Function detail X Y Z Default values
Contact Settings
40 *S 1 Contact S1 Contact 1 inactive 0 2.0.0
Low alarm configuration 1
High alarm configuration 2
Active during HOLD 3
Activation by process value 0
Proportional duty cycle 1
Proportional pulse frequency 2
Activation by temperature 3
PI control inactive 0
PI control active 1
41 * S 2 Contact S2 Contact 2 inactive 0 1.0.0
Low alarm configuration 1
High alarm configuration 2
Active during HOLD 3
Activation by process value 0
Proportional duty cycle 1
Proportional pulse frequency 2
Activation by temperature 3
PI control inactive 0
PI control active 1
42 *S 3 Contact S3 Contact 3 inactive 0 4.0.0
Low alarm configuration 1
High alarm configuration 2
Active during HOLD 3
Active during WASH 4
Activation by process value 0
Proportional duty cycle 1
Proportional pulse frequency 2
Activation by temperature 3
PI control inactive 0
PI control active 1
43 *S 4 Contact S4 Contact 4 inactive 0 4.0.0
Low alarm configuration 1
High alarm configuration 2
Active during HOLD 3
Active during FAIL 4
Activation by process value 0
Proportional duty cycle 1
Proportional pulse frequency 2
Activation by temperature 3
PI control inactive 0
PI control active 1